Initial Rinse
Following clinical use, all instruments must be thoroughly rinsed with cold or lukewarm water immediately to remove bodily fluids, tissue and chemical residues prior to instrument processing.
INSTRUMENT PROCESSING & CARE
Standard operating procedures for cleaning, ultrasonic decontamination, inspection, and autoclave sterilization of MR SHAH CO precision dental and surgical instruments.
Following clinical use, all instruments must be thoroughly rinsed with cold or lukewarm water immediately to remove bodily fluids, tissue and chemical residues prior to instrument processing.
Place instruments into an ultrasonic cleaning bath for a period of 20 minutes to achieve complete particulate removal. Ultrasonic solution should be replaced daily or whenever visibly soiled.
Rinse instruments thoroughly with demineralized water and ensure complete drying before autoclaving. All stainless-steel instruments must be inspected for cleanliness and joint mobility.
Autoclave instruments following standard validated sterilization parameters (134°C / 273°F). Ensure hinged instruments are sterilized in an open position and hinges lubricated with medical-grade lubricant.
Instrument care and processing guidelines are formally published in the original MR SHAH CO product catalogue (physical page 47). These steps ensure longevity, corrosion resistance, and clinical safety.
For specialized questions regarding instrument maintenance, material specifications, or custom reprocessing protocols, please contact our quality team.